Descrizione Clinica
Other
Description
TENS stimulator, EMS and double output channel MASSAGGIO. Before use, read carefully all instructions contained in this user manual and keep it safe for future use.
The Tesmed TE-880R PLUS stimulator belongs to the group of electrical stimulation systems. It has three basic functions: TENS (transcutaneous electric nerve stimulation), EMS (electronic muscle stimulation) and MASSAGGIO.
Functions of the Tesmed TE-880R PLUS stimulator: the device has 73 programs (11 TENS programs, 10 EMS programs and 52 MASSAGE programs) and applies electrical currents in the range of low frequencies for therapy. Each program controls generated electrical impulses, their intensity, frequency and pulse width.
Based on the simulation of the natural impulses of the body, the mechanism of electrical stimulation equipment is to create electrical impulses that are transmitted transcutanely to nerves or muscle fibers through the electrode. The intensity of the two channels can be adjusted independently and applied individually to a part of the body. This double channel device can be used with eight electrodes. The electrical impulse is first transmitted to the tissue and then transmits stimulation to the nerves and muscles in the various parts of the body.
The Massage function is a non-medical function. The Massage stimulation program provides relaxing muscle vibrations.
TENS mode of use:
used for temporary relief of pain associated with sore muscles and neck, shoulders, back, joints, hips, hands, abdomen, upper extremities (arm) and lower extremities (legm) due to exercise stress or normal domestic work.
It is also intended for symptomatic relief and chronic and intractable pain management and pain relief associated with arthritis.
EMS mode:
1) relaxation of muscle spasms;
2) prevention or delay of disuse atrophy;
3) increase in local blood circulation;
4) muscle re-education;
5) immediate post-surgical stimulation of the calf to prevent venous thrombosis;
6) maintenance or increase of the range of TENS or EMS movements can be used at home or in hospital, use must be done by patients aged 18 or older.
Mode of use
1. Connect electrodes to the caves
Insert the cable connector into the electrode connector. Make sure they are connected properly to ensure good performance.
2. Connect electrode cables to the device
Before proceeding with this step, make sure the device is completely off. Hold between your fingers the isolated part of the cable connector and insert it into the socket on the top of the main device. Make sure the electrode cables are inserted correctly. The device has two output sockets controlled by channel A and channel B at the top of the unit. You can unwield to use a channel with a hollow and four electrodes or both channels with two cables and eight electrodes.
3. Electrodes
- Electrodes must be regularly replaced when they begin to lose their adhesiveness. If you are not sure of electrode adhesive properties, order new replacement electrodes. Follow the application procedures indicated on the electrode packaging when using the new spare electrodes, to maintain optimal stimulation and prevent irritation of the skin.
- Place the electrode on the part of the body that needs treatment, according to the instructions of this user manual. Before use, clean the skin and make sure the skin and electrode are well connected.
- Tesmed TE-880R PLUS is a type of OTC stimulator, suitable for domestic use. Use only according to the user manual, place the electrodes in the position where you feel pain. Perform treatment by adapting it to comfortable sensations.
Electrode positioning
You can also use only two electrodes per cavetto (instead of four), in this case connect the two electrodes to a pair of red and black thorns and cover the other pair of red and black hollow spinets with the appropriate lids supplied.
Switching on
Press the button [ON/OFF/M] to turn on the device; the LCD display lights up. The backlight in the absence of operations remains for about 10 seconds after which the display passes in energy saving mode. After pressing the button [ON/OFF/M] you switch to standby mode.
Attention, the device turns off after about 1 minute of inactivity. After first use, in case of re-ignition, the device automatically could show the selected program previously.
Select the treatment mode
Press the button [ON/OFF/M] to select which treatment mode (TENS-EMS-MASS) will be used.
Select the treatment program
Based on your needs, press the button [P] to select the treatment program.
Starting from the 2 program for each type of treatment the processing area will be displayed on the display: a ball will flash on the human figure of the display to indicate the tractable area with the selected program. For example, for the program 2 TENS will flash a ball on the human figure near the shoulder, to indicate precisely the tractable area.
For TENS and EMS programs the default duration of the treatment is 30 minutes, however, you can adjust the duration of the treatment for TENS and EMS programs.
Programs have a default duration of 30 minutes, if you want to customize the duration of the treatment, after selecting the desired program, before giving intensity, press the button [T] (do not hold) to adjust the duration of treatment.
Select the channel for treatment
By default the two channels are selected, on the display two small arrows appear one to the left and one to the right, with the button [+] you increase the intensity in the same way on both channels. Press the button [CH] to select the channel for treatment. The channel selection is useful if you want to set a different level of intensity on the two channels.
Starting treatment by increasing intensity
Make sure the electrodes are placed on the skin and correctly connected. Press the button [+] to increase the intensity of the selected treatment channel.
Adjust the output intensity
Remember to give intensity on both channels. Press the button [+] to increase the output intensity. It will be increased to a higher level after each pressure. The device has 40 levels of intensity. Please adjust the intensity according to the condition in which you feel comfortable. If the stimulation is felt as too strong, you can press the button [-] to reduce intensity at a lower level each time. When the output intensity of both channels decreases to zero, the stimulator will return to standby mode.
Break the treated and turn off the device
Press the button [ON/OFF/M] to stop treatment during treatment mode. Hold the button for 3 seconds [ON/OFF/M] if you want to turn off the electrostimulator and the LCD will turn off.
For MASS programs - how to select the type of massage
Depending on your needs, in MASSAGGIO mode, you can select between 4 different types of massage depending on your user preferences. To select the 4 types of massage press each time for 3 seconds the button [P].
Customizable U programs - how to set parameters
For the time parameter simply press the T button to increase the time of 5 minutes at a time. Press the T button to access the pulse frequency setting mode, pulse amplitude and treatment time. Once you select the parameter you want to change, press the button [+] or [-] to select the desired value. Once you determine the desired parameters press the ON button and give intensity to start the treatment.
Detachable electrode detection
In case of electrodes not perfectly adhered to the skin, perhaps because worn, or lack of connection of the cables to the device, above the level of intensity 5, the device will detect the anomaly by eliminating the intensity and at the same time will flash for about 4 seconds the symbol, 1 or 2 , of the reference channel. Then the stimulator will return to standby mode.
Low battery detection
When the battery is low, the icon will flash to indicate to turn off the device and recharge the battery.
CHARGE THE BATTERY
To recharge the battery, proceed as follows
- This device cannot be used during charging.
- Make sure the device is no longer connected to the patient (output cables and electrodes must be disconnected).
- Connect the micro USB cable to the charging port of the device.
- Connect the USB cable to the charger.
- When the device is in charge, the light indicator is red.
- It could take up to 2 hours to reach the full charge.
- When the charge is completed, the light indicator becomes green.
The life of a rechargeable battery depends on the number of charging/performance cycles to which it is subject and how these cycles are performed.
The following tips help extend battery life:
- When the device is not used frequently, charge the battery once a month.
- For a longer battery life, download it as much as possible.
Use of electrodes
1. The electrode can only be connected with the TE-880R PLUS electrostimulator. Make sure the device is switched off while connecting or disconnecting the electrodes.
2. If you want to reposition the electrode during the application, turn off the device.
3. The use of electrode can cause skin irritation. If such skin irritations occur, e.g. redness, blisters or itching, immediately stop using the device. Do not use Tesmed TE-880R PLUS permanently on the same part of the body, as this may also cause skin irritations.
4. Electrodes are private and intended for the use of a single person. Do not use in multiple people.
5. The electrode must adhere perfectly to the skin to avoid hot spots, which can cause skin burns.
6. It is not recommended to use electrodes when they do not adhere more well to the skin. With proper use on average you can do about 30 treatments.
7. The adhesive force of the electrodes depends on the properties of the skin, storage conditions and the number of applications. If the electrodes are no longer fully attached to the skin surface, replace them with new ones.
Reposition the electrodes on the protective film after use and reposition them in the case to prevent the gel from drying. This maintains the adhesive force for a longer period.
Tips for use for TENS:
1. if the output intensity is too strong, press the button [-] to decrease it;
2. if you do not feel any discomfort during treatment, it is recommended to use the device until the end of the session;
3. pain relief after 5-10 minutes of treatment, we recommend 1-2 treatments per day for at least one week;
4. after a period of treatment, if pain relief is not reached or pain worsens, consult your doctor.
Suggestions for EMS:
1. place the electrodes on the part of the body you want to treat;
2. 1-2 treatments per day, for a variable period;
3. it is recommended to use the device for one session at a time;
4. if you feel discomfort during treatment, you can take a break or decrease the intensity.
Tips for use in MASSAGE:
1. place the electrodes on the part of the body you want to treat;
2. 1-2 treatments per day, for a variable period;
3. if you feel discomfort during treatment, you can take a break or decrease the intensity;
4. massage treatments are designed for people with muscle stiffness, feeling of tension and knotty muscles. The different types of massage provide different sensations, the user can choose the type of favorite massage.
IMPASTATION massage: medium frequency buttons sensations simulate the manipulation with a kneading movement.
SFREGMENT massage: high frequency buttons imitate the rubbing of the hands, simulating the action to press against the muscles, against each other, with a repeated circular movement or up and down.
PERCUSSION massage: medium frequency buttons sensations simulate the effect of acting on the muscles with the knuckles of the hand.
PICCHIETTATION Massage: sensations of low-frequency pickling buttons is as if you slightly hit your muscles by pinching them.
Warnings
- Do not use this device if you use a heart pacemaker, an implanted defibrillator or other metal or electronic implanted devices. Such use could cause electrical shocks, burns, electrical interference or death.
- The device should not be used in the presence of carcinogenic injuries or other injuries in the treatment area.
- Stimulation should not be applied on swollen, infected, inflamed or rashes (e.g. flebite, thrombophlebite, varicose veins, etc.).
- Electrode placements must be avoided in the area of the carotid breast (front neck) or brain (on the head).
- This device should not be used in excessively denerved areas.
- Inguinal hernia.
- Do not use on areas operated after surgery for at least 10 months after surgery.
- Do not use with severe arterial circulatory problems in the lower limbs.
1) If you have had a medical or physical treatment for pain, consult your physician before use.
2) If the pain is not attenuated, or it does not become mild or last more than five days, stop using the device and consult your doctor.
3) Do not apply stimulation on the neck because this could cause severe muscle spasms resulting in the closure of the respiratory tract, difficulty breathing or adverse effects on the heart rhythm or blood pressure.
4) The international rules in force impose the reporting on the application of electrodes at the chest level (increasing the risk of cardiac fibrillation).
5) Do not apply stimulation in the presence of electronic monitoring equipment (e.g. heart monitor, ECG alarms), which may not function properly when the electrical stimulation device is in use.
6) Do not apply stimulation during the bath or shower.
7) Do not apply stimulation during sleep.
8) Do not apply stimulation during driving, working machinery or during any activity where electrical stimulation can lead to the risk of injury.
9) Apply stimulation only on normal, intact, clean and healthy skin.
10) The long-term effects of electrical stimulation are not known. The electrical stimulation device cannot replace the drugs.
11) Stimulation should not occur while the user is connected to high-frequency surgical equipment, which can cause skin burns under electrodes, as well as problems with the stimulator.
12) Do not use the stimulator near short waves or microwave therapy equipment, since this can compromise the output power of the electrostimulator.
13) Never use near the heart area.Stimulation electrodes should never be placed anywhere in the front of the chest (marked by the ribs and sternum), but especially not on the two large pectoral muscles.There, the risk of cardiac fibrillation may increase.
14) Never use on the eye, head and face area.
15) Never use near genitals.
16) Never use on skin areas without normal sensitivity.
17) Keep electrodes separated during treatment. It could cause improper stimulation or skin burns if the electrodes are in contact with each other.
18) Keep the stimulator out of the reach of children.
19) Consult your doctor in case of doubt.
20) Stop the treatment and do not increase the intensity level if you feel discomfort during use.
Precautions
1) TENS is not effective for central pain including headaches.
2) LaTENS does not replace pain medications and other pain management therapies.
3) TENS is a symptomatic treatment and, as such, suppresses the feeling of pain that would otherwise serve as a protective mechanism.
4) Effectiveness can also depend on the selection of a qualified professional in patient management with pain.
5) Since the effects of brain stimulation are unknown, stimulation should not be applied on the head and electrodes should not be placed on opposite sides of the head.
6) The safety of electrical stimulation during pregnancy has not been demonstrated.
7) irritation or skin hypersensitivity may occur due to electrical stimulation or electric conductive medium (silica gel).
8) If you are suspected or diagnosed with heart disease or epilepsy, follow the precautions recommended by your doctor.
9) Be careful if you tend to bleed internally, for example following an injury or fracture.
10) Consult your physician before using your device after a recent surgical procedure, since stimulation can stop the healing process.
11) Consult your physician if you intend to apply stimulation during menstruation or during pregnancy.
12) Only for use of a single patient.
13) This stimulator should not be used by patients who are not suitable or emotionally disturbed, including those with dementia or mental cognitive deficiency.
14) Instructions for use are listed and must be respected; any misuse can be dangerous.
15) Rare cases of skin irritation may occur in the electrode positioning area after a long-term application.
16) Do not use this device in the presence of other devices that send electrical impulses to your body.
17) Do not use pointed objects such as a pencil or ballpoint to operate the buttons on the control panel.
18) Check electrode connections before use.
19) Electric stimulators should only be used with the electrodes recommended by the manufacturer.
Adverse reactions:
1) Possible skin irritations or burns may occur under the electrodes.
2) On very rare occasions, users who use TENS for the first time report having light or faint head. It is recommended to use the product from sitting until you get used to the feeling.
3) If stimulation is uncomfortable, reduce the intensity of stimulation at a comfortable level and contact your doctor if problems persist.
Attention
1. Always remove electrodes from the skin with a slight traction to avoid injury in case of highly sensitive skin.
2. Before applying self-adhesive electrodes, it is recommended to wash and degrease the skin, then dry it well.
3. Do not turn on the device when self-adhesive electrodes are not placed on the body.
4. Remove or move the electrodes by switching off the device first or the appropriate channel so as to avoid involuntary irritations.
5. It is recommended to use self-adhesive electrodes of at least 40x40 mm in the treatment area.
6. Never remove self-adhesive electrodes from the skin while the device is still on.
TENS stimulator, EMS and double output channel MASSAGGIO. Before use, read carefully all instructions contained in this user manual and keep it safe for future use.
The Tesmed TE-880R PLUS stimulator belongs to the group of electrical stimulation systems. It has three basic functions: TENS (transcutaneous electric nerve stimulation), EMS (electronic muscle stimulation) and MASSAGGIO.
Functions of the Tesmed TE-880R PLUS stimulator: the device has 73 programs (11 TENS programs, 10 EMS programs and 52 MASSAGE programs) and applies electrical currents in the range of low frequencies for therapy. Each program controls generated electrical impulses, their intensity, frequency and pulse width.
Based on the simulation of the natural impulses of the body, the mechanism of electrical stimulation equipment is to create electrical impulses that are transmitted transcutanely to nerves or muscle fibers through the electrode. The intensity of the two channels can be adjusted independently and applied individually to a part of the body. This double channel device can be used with eight electrodes. The electrical impulse is first transmitted to the tissue and then transmits stimulation to the nerves and muscles in the various parts of the body.
The Massage function is a non-medical function. The Massage stimulation program provides relaxing muscle vibrations.
TENS mode of use:
used for temporary relief of pain associated with sore muscles and neck, shoulders, back, joints, hips, hands, abdomen, upper extremities (arm) and lower extremities (legm) due to exercise stress or normal domestic work.
It is also intended for symptomatic relief and chronic and intractable pain management and pain relief associated with arthritis.
EMS mode:
1) relaxation of muscle spasms;
2) prevention or delay of disuse atrophy;
3) increase in local blood circulation;
4) muscle re-education;
5) immediate post-surgical stimulation of the calf to prevent venous thrombosis;
6) maintenance or increase of the range of TENS or EMS movements can be used at home or in hospital, use must be done by patients aged 18 or older.
Mode of use
1. Connect electrodes to the caves
Insert the cable connector into the electrode connector. Make sure they are connected properly to ensure good performance.
2. Connect electrode cables to the device
Before proceeding with this step, make sure the device is completely off. Hold between your fingers the isolated part of the cable connector and insert it into the socket on the top of the main device. Make sure the electrode cables are inserted correctly. The device has two output sockets controlled by channel A and channel B at the top of the unit. You can unwield to use a channel with a hollow and four electrodes or both channels with two cables and eight electrodes.
3. Electrodes
- Electrodes must be regularly replaced when they begin to lose their adhesiveness. If you are not sure of electrode adhesive properties, order new replacement electrodes. Follow the application procedures indicated on the electrode packaging when using the new spare electrodes, to maintain optimal stimulation and prevent irritation of the skin.
- Place the electrode on the part of the body that needs treatment, according to the instructions of this user manual. Before use, clean the skin and make sure the skin and electrode are well connected.
- Tesmed TE-880R PLUS is a type of OTC stimulator, suitable for domestic use. Use only according to the user manual, place the electrodes in the position where you feel pain. Perform treatment by adapting it to comfortable sensations.
Electrode positioning
You can also use only two electrodes per cavetto (instead of four), in this case connect the two electrodes to a pair of red and black thorns and cover the other pair of red and black hollow spinets with the appropriate lids supplied.
Switching on
Press the button [ON/OFF/M] to turn on the device; the LCD display lights up. The backlight in the absence of operations remains for about 10 seconds after which the display passes in energy saving mode. After pressing the button [ON/OFF/M] you switch to standby mode.
Attention, the device turns off after about 1 minute of inactivity. After first use, in case of re-ignition, the device automatically could show the selected program previously.
Select the treatment mode
Press the button [ON/OFF/M] to select which treatment mode (TENS-EMS-MASS) will be used.
Select the treatment program
Based on your needs, press the button [P] to select the treatment program.
Starting from the 2 program for each type of treatment the processing area will be displayed on the display: a ball will flash on the human figure of the display to indicate the tractable area with the selected program. For example, for the program 2 TENS will flash a ball on the human figure near the shoulder, to indicate precisely the tractable area.
For TENS and EMS programs the default duration of the treatment is 30 minutes, however, you can adjust the duration of the treatment for TENS and EMS programs.
Programs have a default duration of 30 minutes, if you want to customize the duration of the treatment, after selecting the desired program, before giving intensity, press the button [T] (do not hold) to adjust the duration of treatment.
Select the channel for treatment
By default the two channels are selected, on the display two small arrows appear one to the left and one to the right, with the button [+] you increase the intensity in the same way on both channels. Press the button [CH] to select the channel for treatment. The channel selection is useful if you want to set a different level of intensity on the two channels.
Starting treatment by increasing intensity
Make sure the electrodes are placed on the skin and correctly connected. Press the button [+] to increase the intensity of the selected treatment channel.
Adjust the output intensity
Remember to give intensity on both channels. Press the button [+] to increase the output intensity. It will be increased to a higher level after each pressure. The device has 40 levels of intensity. Please adjust the intensity according to the condition in which you feel comfortable. If the stimulation is felt as too strong, you can press the button [-] to reduce intensity at a lower level each time. When the output intensity of both channels decreases to zero, the stimulator will return to standby mode.
Break the treated and turn off the device
Press the button [ON/OFF/M] to stop treatment during treatment mode. Hold the button for 3 seconds [ON/OFF/M] if you want to turn off the electrostimulator and the LCD will turn off.
For MASS programs - how to select the type of massage
Depending on your needs, in MASSAGGIO mode, you can select between 4 different types of massage depending on your user preferences. To select the 4 types of massage press each time for 3 seconds the button [P].
Customizable U programs - how to set parameters
For the time parameter simply press the T button to increase the time of 5 minutes at a time. Press the T button to access the pulse frequency setting mode, pulse amplitude and treatment time. Once you select the parameter you want to change, press the button [+] or [-] to select the desired value. Once you determine the desired parameters press the ON button and give intensity to start the treatment.
Detachable electrode detection
In case of electrodes not perfectly adhered to the skin, perhaps because worn, or lack of connection of the cables to the device, above the level of intensity 5, the device will detect the anomaly by eliminating the intensity and at the same time will flash for about 4 seconds the symbol, 1 or 2 , of the reference channel. Then the stimulator will return to standby mode.
Low battery detection
When the battery is low, the icon will flash to indicate to turn off the device and recharge the battery.
CHARGE THE BATTERY
To recharge the battery, proceed as follows
- This device cannot be used during charging.
- Make sure the device is no longer connected to the patient (output cables and electrodes must be disconnected).
- Connect the micro USB cable to the charging port of the device.
- Connect the USB cable to the charger.
- When the device is in charge, the light indicator is red.
- It could take up to 2 hours to reach the full charge.
- When the charge is completed, the light indicator becomes green.
The life of a rechargeable battery depends on the number of charging/performance cycles to which it is subject and how these cycles are performed.
The following tips help extend battery life:
- When the device is not used frequently, charge the battery once a month.
- For a longer battery life, download it as much as possible.
Use of electrodes
1. The electrode can only be connected with the TE-880R PLUS electrostimulator. Make sure the device is switched off while connecting or disconnecting the electrodes.
2. If you want to reposition the electrode during the application, turn off the device.
3. The use of electrode can cause skin irritation. If such skin irritations occur, e.g. redness, blisters or itching, immediately stop using the device. Do not use Tesmed TE-880R PLUS permanently on the same part of the body, as this may also cause skin irritations.
4. Electrodes are private and intended for the use of a single person. Do not use in multiple people.
5. The electrode must adhere perfectly to the skin to avoid hot spots, which can cause skin burns.
6. It is not recommended to use electrodes when they do not adhere more well to the skin. With proper use on average you can do about 30 treatments.
7. The adhesive force of the electrodes depends on the properties of the skin, storage conditions and the number of applications. If the electrodes are no longer fully attached to the skin surface, replace them with new ones.
Reposition the electrodes on the protective film after use and reposition them in the case to prevent the gel from drying. This maintains the adhesive force for a longer period.
Tips for use for TENS:
1. if the output intensity is too strong, press the button [-] to decrease it;
2. if you do not feel any discomfort during treatment, it is recommended to use the device until the end of the session;
3. pain relief after 5-10 minutes of treatment, we recommend 1-2 treatments per day for at least one week;
4. after a period of treatment, if pain relief is not reached or pain worsens, consult your doctor.
Suggestions for EMS:
1. place the electrodes on the part of the body you want to treat;
2. 1-2 treatments per day, for a variable period;
3. it is recommended to use the device for one session at a time;
4. if you feel discomfort during treatment, you can take a break or decrease the intensity.
Tips for use in MASSAGE:
1. place the electrodes on the part of the body you want to treat;
2. 1-2 treatments per day, for a variable period;
3. if you feel discomfort during treatment, you can take a break or decrease the intensity;
4. massage treatments are designed for people with muscle stiffness, feeling of tension and knotty muscles. The different types of massage provide different sensations, the user can choose the type of favorite massage.
IMPASTATION massage: medium frequency buttons sensations simulate the manipulation with a kneading movement.
SFREGMENT massage: high frequency buttons imitate the rubbing of the hands, simulating the action to press against the muscles, against each other, with a repeated circular movement or up and down.
PERCUSSION massage: medium frequency buttons sensations simulate the effect of acting on the muscles with the knuckles of the hand.
PICCHIETTATION Massage: sensations of low-frequency pickling buttons is as if you slightly hit your muscles by pinching them.
Warnings
- Do not use this device if you use a heart pacemaker, an implanted defibrillator or other metal or electronic implanted devices. Such use could cause electrical shocks, burns, electrical interference or death.
- The device should not be used in the presence of carcinogenic injuries or other injuries in the treatment area.
- Stimulation should not be applied on swollen, infected, inflamed or rashes (e.g. flebite, thrombophlebite, varicose veins, etc.).
- Electrode placements must be avoided in the area of the carotid breast (front neck) or brain (on the head).
- This device should not be used in excessively denerved areas.
- Inguinal hernia.
- Do not use on areas operated after surgery for at least 10 months after surgery.
- Do not use with severe arterial circulatory problems in the lower limbs.
1) If you have had a medical or physical treatment for pain, consult your physician before use.
2) If the pain is not attenuated, or it does not become mild or last more than five days, stop using the device and consult your doctor.
3) Do not apply stimulation on the neck because this could cause severe muscle spasms resulting in the closure of the respiratory tract, difficulty breathing or adverse effects on the heart rhythm or blood pressure.
4) The international rules in force impose the reporting on the application of electrodes at the chest level (increasing the risk of cardiac fibrillation).
5) Do not apply stimulation in the presence of electronic monitoring equipment (e.g. heart monitor, ECG alarms), which may not function properly when the electrical stimulation device is in use.
6) Do not apply stimulation during the bath or shower.
7) Do not apply stimulation during sleep.
8) Do not apply stimulation during driving, working machinery or during any activity where electrical stimulation can lead to the risk of injury.
9) Apply stimulation only on normal, intact, clean and healthy skin.
10) The long-term effects of electrical stimulation are not known. The electrical stimulation device cannot replace the drugs.
11) Stimulation should not occur while the user is connected to high-frequency surgical equipment, which can cause skin burns under electrodes, as well as problems with the stimulator.
12) Do not use the stimulator near short waves or microwave therapy equipment, since this can compromise the output power of the electrostimulator.
13) Never use near the heart area.Stimulation electrodes should never be placed anywhere in the front of the chest (marked by the ribs and sternum), but especially not on the two large pectoral muscles.There, the risk of cardiac fibrillation may increase.
14) Never use on the eye, head and face area.
15) Never use near genitals.
16) Never use on skin areas without normal sensitivity.
17) Keep electrodes separated during treatment. It could cause improper stimulation or skin burns if the electrodes are in contact with each other.
18) Keep the stimulator out of the reach of children.
19) Consult your doctor in case of doubt.
20) Stop the treatment and do not increase the intensity level if you feel discomfort during use.
Precautions
1) TENS is not effective for central pain including headaches.
2) LaTENS does not replace pain medications and other pain management therapies.
3) TENS is a symptomatic treatment and, as such, suppresses the feeling of pain that would otherwise serve as a protective mechanism.
4) Effectiveness can also depend on the selection of a qualified professional in patient management with pain.
5) Since the effects of brain stimulation are unknown, stimulation should not be applied on the head and electrodes should not be placed on opposite sides of the head.
6) The safety of electrical stimulation during pregnancy has not been demonstrated.
7) irritation or skin hypersensitivity may occur due to electrical stimulation or electric conductive medium (silica gel).
8) If you are suspected or diagnosed with heart disease or epilepsy, follow the precautions recommended by your doctor.
9) Be careful if you tend to bleed internally, for example following an injury or fracture.
10) Consult your physician before using your device after a recent surgical procedure, since stimulation can stop the healing process.
11) Consult your physician if you intend to apply stimulation during menstruation or during pregnancy.
12) Only for use of a single patient.
13) This stimulator should not be used by patients who are not suitable or emotionally disturbed, including those with dementia or mental cognitive deficiency.
14) Instructions for use are listed and must be respected; any misuse can be dangerous.
15) Rare cases of skin irritation may occur in the electrode positioning area after a long-term application.
16) Do not use this device in the presence of other devices that send electrical impulses to your body.
17) Do not use pointed objects such as a pencil or ballpoint to operate the buttons on the control panel.
18) Check electrode connections before use.
19) Electric stimulators should only be used with the electrodes recommended by the manufacturer.
Adverse reactions:
1) Possible skin irritations or burns may occur under the electrodes.
2) On very rare occasions, users who use TENS for the first time report having light or faint head. It is recommended to use the product from sitting until you get used to the feeling.
3) If stimulation is uncomfortable, reduce the intensity of stimulation at a comfortable level and contact your doctor if problems persist.
Attention
1. Always remove electrodes from the skin with a slight traction to avoid injury in case of highly sensitive skin.
2. Before applying self-adhesive electrodes, it is recommended to wash and degrease the skin, then dry it well.
3. Do not turn on the device when self-adhesive electrodes are not placed on the body.
4. Remove or move the electrodes by switching off the device first or the appropriate channel so as to avoid involuntary irritations.
5. It is recommended to use self-adhesive electrodes of at least 40x40 mm in the treatment area.
6. Never remove self-adhesive electrodes from the skin while the device is still on.