Descrizione Clinica
Thermometers - traditional thermometers - temperature devices flex thermometer mt1931 Colpharma
Thermometer Flex MT1931
Medical device CE class IIa.Thermometer measuring in 60 seconds, without mercury.
With flexible probe, memory of the last measurement.
Waterproof with acoustic signal.
Important precautionary measures
- The thermometer is used exclusively for measuring body temperature.
- The thermometer successfully terminates the misu-ration of body temperature only when it is reported acoustically by >.
- The thermometer contains small components (battery, etc.) which can be swallowed by children. Never leave the thermometer for children.
- Protect the thermometer from impacts.
- Avoid tilting the probe over 45°.
- Avoid ambient temperatures above 60 °C: do not boil the thermometer.
- Use only the disinfectants listed in section > to clean the thermome-tro. The thermometer must be intact when immersing it in the disinfecting liquid.
Mode of use
Thermometer ignition:
To turn on the thermometer, press the button next to the display; a short acoustic signal indicates that the thermometer is on. A display test is automatically performed.
All icons must appear on the display.
When the ambient temperature is less than 32 °C it will appear in the display a > (Low – temperature too low to be detected) and at the top right a > flashing. The thermometer is now ready for use.
Instructions for use:
Choose the preferred measurement method.
During the measurement, the temperature is displayed on the display continuously and the > symbol flashes.
When the temperature increase is less than 0.1 °C every 16 seconds the sound signaler (bip-bip-bip-bip-bip for 10 times) will notice that the temperature measurement is finished, the > symbol will cease to flash and the thermometer will be ready for reading the body temperature.
To extend battery life, turn off the thermometer after use briefly pressing the power button and turn off.
The thermometer will automatically turn off after about 10 minutes.
Recording of measured values:
If you hold the button for more than three seconds when you turn on the thermometer, the temperature recorded during the last measurement will automatically appear on the display and a small > will indicate that reading is a data in memory.
After 3 seconds this value will disappear and the thermometer will return to the normal measurement mode.
The recorded value is deleted.
Cod. MT