Descrizione Clinica
Diabetes and blood sugar - Contour next blood sugar 50str Contour
Contour
next
Medical device CE class IIb. CE certification n° LRQ 400 3509/B.Reactive strips for measuring blood sugar to be used with Bayer Contour XT, Contour Next, Contour Next usb and Contour Next Link.
They use a small amount of blood (0.6 μL) which is automatically aspired by capillaries within the reactive ester of the strip: after a 5 seconds countdown, the glucose concentration in the blood sample appears automatically on the viewer. Glucose in the blood sample reacts with the FAD glucose dehydrogenase (FAD-GDH) and the mediator, generating a flow of electrons that produces a current proportional to the concentration of glucose in the sample. The stripes CONTOUR NEXT allow the determination of glucose both on whole capillary blood and on venous blood.
Reagent
Glucose dehydrogenase 21% p/p.
Ombudsman 54% p/p.
Non-reactive substances 25% p/p.
Conservation
At room temperature (0°-30°C).
Do not expose to direct sunlight and moisture.
Do not remove the drying from the packaging.
Format
50 striped bottle.
Cod. 7306/ 7374